Chinese (Simplified)Chinese (Traditional)EnglishFrenchIndonesianJapaneseKoreanRussianThaiKhmerLao

  • Fluke Norma 4000 Power Analyzers

Fluke Norma 4000 Power Analyzers

Specifications: Fluke Norma 4000 Power Analyzers

General Specifications
Number of Phases1 to 3
WeightApprox. 5 kg (11 lb)
Size150 x 237 x 315 mm
5.9 x 9.3 x 12.4 in
On-board PrinterNo
DisplayColor, 5.7 in / 144 mm - 320 x 240 pixel
User-selectable background lighting and contrast.
BandwidthDC to 3 MHz or DC to 10 MHz depending on input module
Basic Accuracy0.2%, 0.1% or 0.03% depending on input modules
Sampling Rate0.33 MHz or 1 MHz depending on input modules
Voltage Input Range0.3 V to 1000 V
Current Input Range (direct, not via shunt)0.03 mA to 20 A depending on input module
Memory for Configurations4 MB
Memory for Settings0.5 MB
Fast Fourier Transformation (FFT)To the 40th harmonic
RS232/USB InterfaceStandard
PI1 Process interface (8 analog/impulse inputs and 4 analog outputs)Optional
IEEE 488.2/GPIB interface (1 MBit/s Ethernet / 10 MBit/s or 100 MBit/s)Optional
Fluke NormaView PC software (for data download, analysis and report writing)Standard
Basic Functions
Fast Fourier Transformation (FFT)Calculation of harmonics with graphical representation. Up to 3 bar graphs are displayed at the same time.
Measured values: U, I, P per phase
Order: 1st to 40th harmonics, maximum half sample frequency
Digital Oscilloscope (DSO)Simultaneous display of up to 3 measured values on sample level. Quick view of curve form and distortion.
Integration function (energy)Simultaneous display of up to 6 configurable numeric values. Start/Stop conditions and positive negative direction available.
Vector DisplayVector display of HO1 up to 6 signals. For easy testing of the right connection of the instrument and quick overview of the phase angle of each signal.
RecorderDisplay of average values over time for trend determination.
RAM data memoryStoring of sample and average values; setting of start and stop conditions.
From the RAM approximately 4 MB are available for the storage of measured values.
ConfigurationSet up the analyzer to measure and display data in the format required.
Ambient Conditions
Working Temperature Range5°C to 35°C (41°F to 95°F)
Storage Temperature Range-20°C to 50°C (-4°F to 122°F)
Housing MaterialFluke Norma Power Analyzers are extremely compact and equipped with a solid metal case to meet stringent EMC requirements.
Climatic ClassKYG DIN 40040, max. 85% relative humidity, non-condensing.
Power supply85 V AC to 264 V AC, 50 Hz to 60 Hz, DC 100 to 260 V, ca. 40 VA European plug with switch. Binding post for current available on some models.
Measuring InputsSafety sockets 4 mm, 2 for each input. External shunt connection over BNC socket.
OperationMembrane keyboard with cursor – function keys and direct functions.
ConnectionsRear panel of the 3-phase Analyzer
Measured Values
 

Non-gapping calculation of averaged values for each phase. In three phase system additionally calculation of total power and averaging of V and I of the three phases. The fundamental H01 will be calculated in synchronous mode also for these values.

  • Urms effective value, Urm rectified mean, Um mean value
  • Up-, Up+, Upp peak values
  • Ucf crest factor Ucf, Uff form factor
  • Ufc fundamental content
  • Uthd distortion factor DIN, IEC
  • Irms effective value, Irm rectified mean, Im mean value
  • Ip-, Ip+, Ipp peak values
  • Icf crest factor Icf, Iff form factor
  • Ifc fundamental content
  • Ithd distortion factor DIN, IEC
  • P active power [W]
  • Q reactive power [Var]
  • S apparent power [VA]
  • ë, cos. phase angular
  • Integral function for active power P, reactive power Q, apparent power S, voltage (Um) and current (Im),
  • Number of digits 4 or 5 dependent on measured value.
Frequency and Synchronization
RangeDC and 0.2 Hz to sample rate
Accuracy±0.01% of measured value (reading)
 
  • Channels which can be selected: all U/I or external input.
  • One of three low pass filter with different frequencies can be switched into the signal.
  • The frequency is always visible on the top of the screen.
  • The BNC synchronization socket on backside of the instrument can be used either as input or output.
  • The input signals can be measured up to the sample rate of the power phase. The maximum level must not be higher then 50 V.
  • The output signal is a pulsed 5Volts TTL signal (frequency depends on the measured synch frequency).
Configuration Memory
 Up to 15 user configurations can be saved into a permanent memory and reloaded later on. Changes that were not saved are lost after switching off the instrument.
Interface
 RS232 interface for upload of firmware and data exchange with the PC. A printer can be connected over an external converter.
OptionsIEEE 488.2 / 1 MBit/s
Ethernet / 10 MBit/s or 100 Mbit/s
Standards and Safety
Electrical SafetyEN 61010-1 / 2nd Edition 1000 V CAT II (600V CAT III)
Degree of pollution 2, safety Class I
EN 61558 for transformer
EN 61010-2-031/032 for accessories
Maximum inputsFor voltage inputs Measurement range 1000 Veff, 2 kVpeak
For current inputs Measurement range 10 Aeff, 20 Apeak
Test voltageNet inputcase (protective conductor): 1.5 KV AC
Net connectionMeasurement input: 5.4 kV AC
Measurement inputsCase: 3.3 kV AC
Measurement inputinput: 5.4 kV
Electromagnetic susceptibilityEmissionIEC 61326-1, EN 50081-1, EN 55011 Class B
ImmunityIEC 61326-1 / Annex A (industrial sector), EN 50082-1

Power Phases

The Fluke Norma 4000 Power Analyzer can be equipped with up to three power phases. Users can select the power phase best suited for their application.

Each modular plug-in power phase consists of a voltage and a current measurement channel. Each measuring channel is available for each basic unit, however only one kind of channel can be used per unit. Check standard configurations.

Power Phase Overview

Power Phase Channels
PP42Accuracy0.2% (0.1% rd + 0.1% rg)
Current range20 A
Sampling rate341 kHz
Bandwidth3 MHz
PP50Accuracy0.1% (0.05% rd + 0.05% rg)
Current range10 A
Sampling rate1 MHz
Bandwidth10 MHz
PP54Accuracy0.1% (0.05% rd + 0.05% rg)
Current range10 A
Sampling rate341 kHz
Bandwidth3 MHz
PP64Accuracy0.03% (0.02% rg + 0.01% rg)
Current range10 A
Sampling rate341 kHz
Bandwidth3 MHz

Current Shunts

Planarshunt
Continuous load (Imin)
Range (Imax)
0.3 A
32 A
Nominal current (calibration point)32 A
Basic accuracy at calibration point [%]± 0.03
Basic accuracy at nominal current [%]± 0.03
Nominal voltage drop [V]320 mV
Nominal resistance [Ohm]10 mΩ
Short time overrange
(5s load 15s interval)
100A/ 1s-5s
Overload Wmax-
BandwidthDC - 1 MHz
Frequency range-
Angular accuracy [°/kHz]± 0.1
Frequency influence [%/kHz]-
Load influence [%/A2]-
Temperature coefficient [ppm/K]≤10
Weight (kg / lbs)0.62 /1.36

Write a review

Note: HTML is not translated!
   Bad           Good

Fluke Norma 4000 Power Analyzers

  • Product Code:Fluke Norma 4000 Power Analyzers
  • Availability:In Stock

Tags: FLUKE

K & P Intertrade Co., Ltd. - Certify

กรมพัฒนาธุรกิจการค้า Trustmarkthai

สินค้าของเรา

เครื่องกำเนิดไฟฟ้า · รอกยกของ · ไม้ชักฟิวส์ · ไม้ฉนวน · บันไดสไลด์ · มีดปอกสาย · ชุดประกอบฉนวนค้ำยันสายไฟ · Ratchet Hoist แบบผ้าใบ · Ratchet Cable Cutter · Insulator Cover · InsulatedHanger · Conductor Cutter with Insulated Handles · Conductor Cover 46 kv · Clamp stick grip all type ขนาด 8 ฟุต · Clamp Stick grip all type ขนาด 12 ฟุต · สว่านไร้สาย · ล้อวัดระยะทาง · ประแจบล๊อคไฟฟ้าไร้สาย (BRUSHLESS CORDLESS IMPACT WRENCH) · เทปวัดระยะชนิดโลหะ · ตู้แผงสวิตช์ · ตู้เก็บเครื่องมือ · เครื่องอัดลมใช้มอเตอร์ไฟฟ้า · เครื่องมือบีบหลอดต่อสายไฮดรอลิค · เครื่องมือบีบคอนเนคเตอร์ แบบ H · เครื่องมือต่อสายลงดิน (Grounding Equipment) · เครื่องมือจับและดึงสาย (Wire Grip Come Along) Overall outside · เครื่องดึงสาย (Coffing Hoist) · อุปกรณ์ระบบไฟฟ้า · สวิตซ์ตัดตอนแรงสูง · หางปลา · หลอดต่อสาย · สลักเข้าปลายสาย · พีจีคอนเนคเตอร์ · ฉนวนป้องกันไฟฟ้าลัดวงจร · ฉนวนครอบล่อฟ้า · เครื่องมือช่างไฟฟ้าอุตสาหกรรม · เครื่องมือไฮดรอลิก · เครื่องมือแบตเตอรี่ไร้สายดิสคอนเนคติ้งสวิตซ์แบริเออร์ · ฉนวนครอบบุชชิ่ง · ตาข่าย · แผ่นอลูมิเนียม · ฉนวนครอบดรอปเอาท์ฟิวส์ · หนามป้องกันสัตว์ · อุปกรณ์ระบบไฟฟ้า · หม้อแปรงไฟฟ้า · ROADMAPTOSTRONGGRIDLV · แคล้มป์สติ๊ก · ClampStick · เชือกใยยักษ์ · เชือก Handline · KNIPEX · เครื่องตรวจจับแรงดันไฟฟ้า · UNI · สายกันตก · UNI9103 · เข็มขัดปีนเสา · YQK · SME · ขายดี · หมวกปีกรอบ · หมวกนิรภัย S-Guard · MSA V-Gard full brim · คีมตัดสาย · คัทเตอร์ตัดสาย · CableCutter · คีมตัดลวดเหล็ก · คัทเตอร์เฟืองทด · คัทเตอร์ตัดสายเคเบิ้ล · คีมย้ำ · คีมบีบ · แคล้มป์สติ๊ก · KANDP · Milwaukee · CHANCE · RITZ · KUDOS · ZUPPER  · FLUKE · Klauke Hydraulic Crimp Tool · Google · 




Official Facebook Page

เนื้อหาส่วนนี้อาจไม่แสดงกรณีใช้ผ่านมือถือบางรุ่น ทางบริษัทฯ ขออภัยในความไม่สะดวก

4 TOP Facebook Official Page Post

เนื้อหาส่วนนี้อาจไม่แสดงกรณีใช้ผ่านมือถือบางรุ่น ทางบริษัทฯ ขออภัยในความไม่สะดวก


Chinese (Simplified)Chinese (Traditional)EnglishFrenchIndonesianJapaneseKoreanRussianThaiKhmerLao