Fluke Norma 4000 Power Analyzers
Specifications: Fluke Norma 4000 Power Analyzers
General Specifications | ||
Number of Phases | 1 to 3 | |
Weight | Approx. 5 kg (11 lb) | |
Size | 150 x 237 x 315 mm | |
5.9 x 9.3 x 12.4 in | ||
On-board Printer | No | |
Display | Color, 5.7 in / 144 mm - 320 x 240 pixel | |
User-selectable background lighting and contrast. | ||
Bandwidth | DC to 3 MHz or DC to 10 MHz depending on input module | |
Basic Accuracy | 0.2%, 0.1% or 0.03% depending on input modules | |
Sampling Rate | 0.33 MHz or 1 MHz depending on input modules | |
Voltage Input Range | 0.3 V to 1000 V | |
Current Input Range (direct, not via shunt) | 0.03 mA to 20 A depending on input module | |
Memory for Configurations | 4 MB | |
Memory for Settings | 0.5 MB | |
Fast Fourier Transformation (FFT) | To the 40th harmonic | |
RS232/USB Interface | Standard | |
PI1 Process interface (8 analog/impulse inputs and 4 analog outputs) | Optional | |
IEEE 488.2/GPIB interface (1 MBit/s Ethernet / 10 MBit/s or 100 MBit/s) | Optional | |
Fluke NormaView PC software (for data download, analysis and report writing) | Standard | |
Basic Functions | ||
Fast Fourier Transformation (FFT) | Calculation of harmonics with graphical representation. Up to 3 bar graphs are displayed at the same time. | |
Measured values: U, I, P per phase | ||
Order: 1st to 40th harmonics, maximum half sample frequency | ||
Digital Oscilloscope (DSO) | Simultaneous display of up to 3 measured values on sample level. Quick view of curve form and distortion. | |
Integration function (energy) | Simultaneous display of up to 6 configurable numeric values. Start/Stop conditions and positive negative direction available. | |
Vector Display | Vector display of HO1 up to 6 signals. For easy testing of the right connection of the instrument and quick overview of the phase angle of each signal. | |
Recorder | Display of average values over time for trend determination. | |
RAM data memory | Storing of sample and average values; setting of start and stop conditions. | |
From the RAM approximately 4 MB are available for the storage of measured values. | ||
Configuration | Set up the analyzer to measure and display data in the format required. | |
Ambient Conditions | ||
Working Temperature Range | 5°C to 35°C (41°F to 95°F) | |
Storage Temperature Range | -20°C to 50°C (-4°F to 122°F) | |
Housing Material | Fluke Norma Power Analyzers are extremely compact and equipped with a solid metal case to meet stringent EMC requirements. | |
Climatic Class | KYG DIN 40040, max. 85% relative humidity, non-condensing. | |
Power supply | 85 V AC to 264 V AC, 50 Hz to 60 Hz, DC 100 to 260 V, ca. 40 VA European plug with switch. Binding post for current available on some models. | |
Measuring Inputs | Safety sockets 4 mm, 2 for each input. External shunt connection over BNC socket. | |
Operation | Membrane keyboard with cursor – function keys and direct functions. | |
Connections | Rear panel of the 3-phase Analyzer | |
Measured Values | ||
Non-gapping calculation of averaged values for each phase. In three phase system additionally calculation of total power and averaging of V and I of the three phases. The fundamental H01 will be calculated in synchronous mode also for these values.
| ||
Frequency and Synchronization | ||
Range | DC and 0.2 Hz to sample rate | |
Accuracy | ±0.01% of measured value (reading) | |
| ||
Configuration Memory | ||
Up to 15 user configurations can be saved into a permanent memory and reloaded later on. Changes that were not saved are lost after switching off the instrument. | ||
Interface | ||
RS232 interface for upload of firmware and data exchange with the PC. A printer can be connected over an external converter. | ||
Options | IEEE 488.2 / 1 MBit/s | |
Ethernet / 10 MBit/s or 100 Mbit/s | ||
Standards and Safety | ||
Electrical Safety | EN 61010-1 / 2nd Edition 1000 V CAT II (600V CAT III) | |
Degree of pollution 2, safety Class I | ||
EN 61558 for transformer | ||
EN 61010-2-031/032 for accessories | ||
Maximum inputs | For voltage inputs Measurement range 1000 Veff, 2 kVpeak | |
For current inputs Measurement range 10 Aeff, 20 Apeak | ||
Test voltage | Net input | case (protective conductor): 1.5 KV AC |
Net connection | Measurement input: 5.4 kV AC | |
Measurement inputs | Case: 3.3 kV AC | |
Measurement input | input: 5.4 kV | |
Electromagnetic susceptibility | Emission | IEC 61326-1, EN 50081-1, EN 55011 Class B |
Immunity | IEC 61326-1 / Annex A (industrial sector), EN 50082-1 |
Power Phases
The Fluke Norma 4000 Power Analyzer can be equipped with up to three power phases. Users can select the power phase best suited for their application.
Each modular plug-in power phase consists of a voltage and a current measurement channel. Each measuring channel is available for each basic unit, however only one kind of channel can be used per unit. Check standard configurations.
Power Phase Overview
Power Phase Channels | ||
PP42 | Accuracy | 0.2% (0.1% rd + 0.1% rg) |
Current range | 20 A | |
Sampling rate | 341 kHz | |
Bandwidth | 3 MHz | |
PP50 | Accuracy | 0.1% (0.05% rd + 0.05% rg) |
Current range | 10 A | |
Sampling rate | 1 MHz | |
Bandwidth | 10 MHz | |
PP54 | Accuracy | 0.1% (0.05% rd + 0.05% rg) |
Current range | 10 A | |
Sampling rate | 341 kHz | |
Bandwidth | 3 MHz | |
PP64 | Accuracy | 0.03% (0.02% rg + 0.01% rg) |
Current range | 10 A | |
Sampling rate | 341 kHz | |
Bandwidth | 3 MHz |
Current Shunts
Planarshunt | |
Continuous load (Imin) Range (Imax) | 0.3 A |
32 A | |
Nominal current (calibration point) | 32 A |
Basic accuracy at calibration point [%] | ± 0.03 |
Basic accuracy at nominal current [%] | ± 0.03 |
Nominal voltage drop [V] | 320 mV |
Nominal resistance [Ohm] | 10 mΩ |
Short time overrange (5s load 15s interval) | 100A/ 1s-5s |
Overload Wmax | - |
Bandwidth | DC - 1 MHz |
Frequency range | - |
Angular accuracy [°/kHz] | ± 0.1 |
Frequency influence [%/kHz] | - |
Load influence [%/A2] | - |
Temperature coefficient [ppm/K] | ≤10 |
Weight (kg / lbs) | 0.62 /1.36 |
Fluke Norma 4000 Power Analyzers
- Product Code:Fluke Norma 4000 Power Analyzers
- Availability:In Stock
Tags: FLUKE
สินค้าของเรา
เครื่องกำเนิดไฟฟ้า ·
รอกยกของ · ไม้ชักฟิวส์ · ไม้ฉนวน · บันไดสไลด์ · มีดปอกสาย
· ชุดประกอบฉนวนค้ำยันสายไฟ · Ratchet Hoist แบบผ้าใบ · Ratchet Cable Cutter · Insulator Cover · InsulatedHanger
· Conductor Cutter with Insulated Handles · Conductor Cover 46 kv · Clamp stick grip
all type ขนาด 8 ฟุต · Clamp Stick
grip all type ขนาด 12 ฟุต · สว่านไร้สาย · ล้อวัดระยะทาง · ประแจบล๊อคไฟฟ้าไร้สาย (BRUSHLESS CORDLESS IMPACT WRENCH) · เทปวัดระยะชนิดโลหะ · ตู้แผงสวิตช์ · ตู้เก็บเครื่องมือ · เครื่องอัดลมใช้มอเตอร์ไฟฟ้า ·
เครื่องมือบีบหลอดต่อสายไฮดรอลิค · เครื่องมือบีบคอนเนคเตอร์
แบบ H · เครื่องมือต่อสายลงดิน (Grounding Equipment)
· เครื่องมือจับและดึงสาย (Wire Grip Come Along) Overall
outside · เครื่องดึงสาย (Coffing Hoist) · อุปกรณ์ระบบไฟฟ้า
· สวิตซ์ตัดตอนแรงสูง · หางปลา · หลอดต่อสาย · สลักเข้าปลายสาย · พีจีคอนเนคเตอร์ · ฉนวนป้องกันไฟฟ้าลัดวงจร ·
ฉนวนครอบล่อฟ้า · เครื่องมือช่างไฟฟ้าอุตสาหกรรม · เครื่องมือไฮดรอลิก · เครื่องมือแบตเตอรี่ไร้สายดิสคอนเนคติ้งสวิตซ์แบริเออร์
· ฉนวนครอบบุชชิ่ง · ตาข่าย · แผ่นอลูมิเนียม · ฉนวนครอบดรอปเอาท์ฟิวส์ · หนามป้องกันสัตว์ · อุปกรณ์ระบบไฟฟ้า · หม้อแปรงไฟฟ้า · ROADMAPTOSTRONGGRIDLV · แคล้มป์สติ๊ก
· ClampStick · เชือกใยยักษ์ · เชือก Handline
· KNIPEX · เครื่องตรวจจับแรงดันไฟฟ้า · UNI · สายกันตก · UNI9103 · เข็มขัดปีนเสา · YQK ·
SME · ขายดี · หมวกปีกรอบ · หมวกนิรภัย S-Guard · MSA V-Gard full brim · คีมตัดสาย · คัทเตอร์ตัดสาย · CableCutter · คีมตัดลวดเหล็ก ·
คัทเตอร์เฟืองทด · คัทเตอร์ตัดสายเคเบิ้ล ·
คีมย้ำ · คีมบีบ · แคล้มป์สติ๊ก
·
KANDP
Official Facebook Page
เนื้อหาส่วนนี้อาจไม่แสดงกรณีใช้ผ่านมือถือบางรุ่น ทางบริษัทฯ ขออภัยในความไม่สะดวก
4 TOP Facebook Official Page Post
เนื้อหาส่วนนี้อาจไม่แสดงกรณีใช้ผ่านมือถือบางรุ่น ทางบริษัทฯ ขออภัยในความไม่สะดวก