Fluke Calibration 6531, 6532 E-DWT Electronic Deadweight Tester Kits
Specifications: Fluke Calibration 6531, 6532 E-DWT Electronic Deadweight Tester Kits
| General | |
| Power requirements | |
| To RPM4-E-DWT: | 12 V dc 1.2 A |
| To AC to DC power supply: | 100 V ac to 240 V ac, 50/60 Hz |
| Temperature range | |
| Storage: | - 20 °C to 70 °C |
| Operating: | 10 °C to 40 °C |
| Relative Humidity | |
| Storage: | 0 % to 100 % |
| Operating: | 0 % to 70 % |
| Weight | |
| 1 Q-RPT: | 12 kg (26 lb) approximate |
| 2 Q-RPT: | 14 kg (30 lb) approximate |
| Dimensions | |
| E-DWT footprint: | 41.4 cm W x 37.1 cm D (16.3 in x 14.6 in) |
| E-DWT height: | 26.9 cm (10.6 in), 33.6 cm (13.2 in) to max variable volume handle height |
| Pressure range | Up to 200 MPa (30k psi), depending on configuration |
| Operating medium | All 6531 and 6532 E-DWT Kits are shipped dry, standard preparation. Can be filled with di-ethyl-hexyl sebacate, silicon oils, propylene glycol, fully fluorinated liquids, partially fluorinated liquids, isopropyl alcohol and distilled water or mineral oil. |
| Reservoir capacity | 300 cc (18 in3) |
| Variable volume displacement | 3 cc (0.18 in3), 200 MPa (30,000 psi) maximum |
| Filling and priming pump displacement | 3.7 cc (0.23 in3) |
| TEST pressure connection | DH500 female, 200 MPa (30k psi) maximum working pressure. Adapters included for 1/8 in, 1/4 in, 3/8 in and 1/2 in NPT, 1/8 in, 1/4 in, 3/8 in and 1/2 in BSP, M20 x 1.5 and M14 x 1.5. Adapter maximum working pressure is 140 MPa (20k psi). Note: DH500 is a gland and collar type fitting for 6 mm (1/4 in) coned and left hand threaded tubes equivalent to AE F250C, HIP HF4, 9/16-18 UNF, etc. |
| Pressure Limits | |
| Maximum Working Pressure: | Range of RPM4-E-DWT monitor’s Hi Q-RPT |
| Maximum priming pump pressure: | 700 kPa (100 psi) |
| Maximum Working Pressure: | with Lo Q-RPT selected: Range of RPM4-E-DWT monitor’s Lo Q-RPT |
| Communication ports: | RS-232 (COM1, COM2) |
| Pressure measurement | |
| Warm up time | 15 minute temperature stabilization recommended from cold power up |
| Normal Operating Temperature Range | 10 °C to 40 °C (50 °F to 104 °F) |
| Resolution | Default: 0.01 % of active range User adjustable to 1 ppm of Q-RPT maximum or 10 ppm of active AutoRange, whichever is larger |
| Precision1 | 200 MPa (30,000 psi) ranges: ± 0.018 % of reading or 0.0036 % of Q-RPT span, whichever is greater All other ranges: ± 0.018 % of reading or 0.0018 % of Q-RPT span, whichever is greater |
| Predicted Stability2 | |
| One year: | ± 0.0075 % of reading |
| Two year: | ± 0.015 % of reading |
| Measurement Uncertainty3 | |
| One year: | 200 MPa (30,000 psi) ranges: ± 0.02 % of reading or 0.004 % of Q-RPT span, whichever is greater All other ranges: ± 0.02 % of reading or 0.002 % of Q-RPT span, whichever is greater |
| Two year: | 200 MPa (30,000 psi) ranges: ± 0.025 % of reading or 0.005 % of Q-RPT span, whichever is greater All other ranges: ± 0.025 % of reading or 0.0025 % of Q-RPT span, whichever is greater |
1. Combined linearity, hysteresis, and repeatability. Precision does not include stability or calibration reference uncertainty.
2. Predicted Q-RPT measurement stability limit (k=2) assuming regular use of AutoZero function and short term stability between rezeroing.
3. Maximum deviation of the Q-RPT indication from the true value of applied pressure including precision, predicted stability with rezeroing, temperature effect from 10 ºC to 40 ºC and calibration uncertainty (assumes calibration reference uncertainty of ±0.005 % of reading, k=2), combined and expanded (k=2) following the ISO “Guide to the Expression of Uncertainty in Measurement.
Fluke Calibration 6531, 6532 E-DWT Electronic Deadweight Tester Kits
- Product Code:Fluke Calibration 6531, 6532
- Availability:In Stock
-