Fluke Calibration E-DWT Electronic Deadweight Tester
Specifications: Fluke Calibration E-DWT Electronic Deadweight Tester
| General | ||
| Power requirements | To RPM4-E-DWT: 12 V DC 1.2 A | |
| To AC to DC power supply: 100 V AC to 240 V AC, 50/60 Hz | ||
| Temperature range | Storage: - 20°C to 70°C | |
| Operating: 10°C to 40°C | ||
| Relative humidity | Storage: 0% to 100% | |
| Operating: 0% to 70% | ||
| Weight | 1 Q-RPT: 12 kg (26 lb) approximate | |
| 2 Q-RPT: 14 kg (30 lb) approximate | ||
| Dimensions | E-DWT footprint (W x D): 41.4 x 37.1 cm (16.3 x 14.6 in) | |
| E-DWT height: 26.9 cm (10.6 in), 33.6 cm (13.2 in) to max variable volume handle height | ||
| Pressure ranges dependent on Q-RPT(s) included in RPM4-E-DWT. | 200 MPa (30,000 psi) maximum with standard variable volume | |
| 100 MPa (15,000 psi) maximum with high volume (-HV) variable volume | ||
| Operating medium | Delivered filled with oil (di-ethyl-hexyl sebacate) or dry | |
| Standard E-DWT-H compatible with Sebacate, silicon oils, propylene glycol, fully fluorinated liquids, partially fluorinated liquids, isopropyl alcohol , and distilled water. Option for Skydrol or mineral oil preparation. | ||
| Reservoir capacity | 300 cc (18 in3) | |
| Variable volume displacement | Standard: 3 cc (0.18 in3), 200 MPa (30,000 psi) maximum | |
| High: 7 cc (0.43 in3), 100 MPa (15,000 psi) maximum | ||
| Filling and priming pump displacement | 3.7 cc (0.23 in3) | |
| TEST pressure connection | DH500 female. | |
| DH500 is a gland and collar type fitting for 6mm (1/4 in) coned and left hand threaded tubes equivalent to AE F250C, HIP HF4, 9/16-18 UNF, etc. | ||
| Pressure limits | Maximum working pressure: | Range of RPM4-E-DWT monitor’s Hi Q-RPT 200 MPa (30,000 psi) with standard variable volume |
| 100 MPa (15,000 psi) with high volume variable volume | ||
| Maximum priming pump pressure: | 700 kPa (100 psi) | |
| Maximum working pressure: | With Lo Q-RPT selected: Range of RPM4-E-DWT monitor’s Lo Q-RPT | |
| Communication ports | RS232 (COM1, COM2) | |
| Pressure Measurement | ||
| Warm up time | 15 minute temperature stabilization recommended from cold power up. | |
| Normal operating temperature range | 10°C to 40°C | |
| Resolution | Default: 0.01% of active range | |
| User adjustable to 1 ppm of Q-RPT maximum or 10 ppm of active AutoRange, whichever is larger | ||
| Precision1 | ±0.018% of reading or 0.0018% of Q-RPT span, whichever is greater | |
| Predicted stability2 | One year: ±0.0075% of reading | |
| Two year: ±0.015% of reading | ||
| Measurement uncertainty3 | One year: ±0.02% of reading or 0.002% of Q-RPT span, whichever is greater | |
| Two year: ±0.025% of reading or 0.0025% of Q-RPT span, whichever is greater | ||
| 1. Combined linearity, hysteresis, and repeatability. Precision does not include stability or calibration reference uncertainty. 2. Predicted Q-RPT measurement stability limit (k=2) assuming regular use of AutoZero function and short term stability between rezeroing. 3. Maximum deviation of the Q-RPT indication from the true value of applied pressure including precision, predicted stability with rezeroing, temperature effect from 10°C to 40°C and calibration uncertainty (assumes calibration reference uncertainty of ±0.005% of reading, k=2), combined and expanded (k=2) following the ISO Guide to the Expression of Uncertainty in Measurement. | ||
Fluke Calibration E-DWT Electronic Deadweight Tester
- Product Code:Fluke Calibration E-DWT Electronic Deadweight Tester
- Availability:In Stock
-
0 บาท
Tags: FLUKE

Official Facebook Page
เนื้อหาส่วนนี้อาจไม่แสดงกรณีใช้ผ่านมือถือบางรุ่น ทางบริษัทฯ ขออภัยในความไม่สะดวก
แคตตาล็อก 2569
บริษัท ฯ ไม่อนุญาตให้นำภาพไปใช้ในทุกกรณี