Fluke 753 Documenting Process Calibrator
Specifications: Fluke 753 Documenting Process Calibrator
| Measurement Accuracy | ||||||||
| Voltage DC | Range /Resolution | 1 Year | 2 Years | |||||
| 100.000 mV | 0.02% + 0.005 mV | 0.03% + 0.005 mV | ||||||
| 3.00000 V | 0.02% + 0.00005 V | 0.03% + 0.00005 V | ||||||
| 30.0000 V | 0.02% + 0.0005 V | 0.03% + 0.0005 V | ||||||
| 300.00 V | 0.05% + 0.05 V | 0.07% + 0.05 V | ||||||
| Voltage AC | 3.000 V (40 Hz to 500 Hz) / 0.001 V | 0.5% + 0.002 V | 1.0% + 0.004 V | |||||
| 30.00 V (40 Hz to 500 Hz) / 0.01 V | 0.5% + 0.02 V | 1.0% + 0.04 V | ||||||
| 300.0 V (40 Hz to 500 Hz) / 0.1 V | 0.5% + 0.2 V | 1.0% + 0.2 V | ||||||
| Current DC | 30.000 mA | 0.01% + 5 uA | 0.015% + 7 uA | |||||
| 110.00 mA | 0.01% + 20 uA | 0.015% + 30 uA | ||||||
| Resistance | 10.000 Ω | 0.05% + 50 mΩ | 0.07% + 70 mΩ | |||||
| 100.00 Ω | 0.05% + 50 mΩ | 0.07% + 70 mΩ | ||||||
| 1.0000 kΩ | 0.05% + 500 mΩ | 0.07% + 0.5 Ω | ||||||
| 10.000 kΩ | 0.1% + 10 Ω | 0.15% + 15 Ω | ||||||
| Frequency | 1.00 to 110.00 Hz / 0.01 Hz | 0.05 Hz | ||||||
| 110.1 to 1100.0 Hz / 0.1 Hz | 0.5 Hz | |||||||
| 1.101 to 11.000 kHz / 0.001 kHz | 0.005 kHz | |||||||
| 11.01 to 50.00 kHz / 0.01 kHz | 0.05 kHz | |||||||
| Source Accuracy | ||||||||
| 1 Year | 2 Years | |||||||
| Voltage DC | 100.000 mV | 0.01% + 0.005 mV | 0.015% + 0.005 mV | |||||
| 1.00000 V | 0.01% + 0.00005 V | 0.015% + 0.0005 V | ||||||
| 15.0000 V | 0.01% + 0.0005 V | 0.015% + 0.0005 V | ||||||
| Current DC | 22.000 mA (source) | 0.01% + 0.003 mA | 0.02% + 0.003 mA | |||||
| Current sink (simulate) | 0.02% + 0.007 mA | 0.04% + 0.007 mA | ||||||
| Resistance | 10.000 Ω | 0.01% + 10 mΩ | 0.015% + 15 mΩ | |||||
| 100.00 Ω | 0.01% + 20 mΩ | 0.015% + 30 mΩ | ||||||
| 1.0000 kΩ | 0.02% + 0.2 Ω | 0.03% + 0.3 Ω | ||||||
| 10.000 kΩ | 0.02% + 3 Ω | 0.03% + 5 Ω | ||||||
| Frequency | 0.1 to 10.99 Hz | 0.01 Hz | ||||||
| 0.01 to 10.99 Hz | 0.01 Hz | |||||||
| 11.00 to 109.99 Hz | 0.1 Hz | |||||||
| 110.0 to 1099.9 Hz | 0.1 Hz | |||||||
| 1.100 to 21.999 kHz | 0.002 kHz | |||||||
| 22.000 to 50.000 kHz | 0.005 kHz | |||||||
| Technical Data | ||||||||
| Data log functions | Measure functions | Voltage, current, resistance, frequency, temperature, pressure | ||||||
| Reading rate | 1, 2, 5, 10, 20, 30, or 60 readings/minute | |||||||
| Maximum record length | 8000 readings (7980 for 30 or 60 readings/minute) | |||||||
| Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature | ||||||
| Rate | 4 steps/second | |||||||
| Trip detect | Continuity or voltage (continuity detection not available when sourcing current) | |||||||
| Loop power function | Voltage | Selectable, 26 V | ||||||
| Accuracy | 10%, 18 V minimum at 22 mA | |||||||
| Maximum current | 25 mA, short circuit protected | |||||||
| Maximum input voltage | 50 V DC | |||||||
| Step functions | Source functions | Voltage, current, resistance, frequency, temperature | ||||||
| Manual step | Selectable step, change with arrow buttons | |||||||
| Autostep | Fully programmable for function, start delay, stepvalue, time per step, repeat | |||||||
| Environmental Specifications | ||||||||
| Operating temperature | -10°C to +50°C | |||||||
| Storage temperature | -20°C to +60°C | |||||||
| Dust/water resistance | Meets IP52, IEC 529 | |||||||
| Operating altitude | 3000 m above mean sea level (9842 ft) | |||||||
| Safety Specifications | ||||||||
| Agency approvals | CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 | |||||||
| Mechanical and General Specifications | ||||||||
| Size | 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in) | |||||||
| Weight | 1.2 kg (2.7 lb) | |||||||
| Batteries | Internal Battery Pack Li-ion: 7.2 V, 4400 mAh, 30 Wh | |||||||
| Battery life | > 8 hours typical | |||||||
| Battery replacement | Via snap-shut door without opening calibrator; no tools required | |||||||
| Side port connections | Pressure module connector | |||||||
| USB Connector to interface to your PC | ||||||||
| Connection for optional battery charger/eliminator | ||||||||
| Data storage capacity | 1 week of calibration procedures results | |||||||
| 90 day specifications | The standard specification interval for the 750 Series are 1 and 2 years. | |||||||
| Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2. | ||||||||
| Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant. | ||||||||
| Temperature, Resistance Temperature Detectors | ||||||||
| Degrees or % of reading - Type (α) | Range °C | Measure °C1 | ||||||
| 1 year | 2 year | |||||||
| 100 Ω Pt (385) | -200 to 100 100 to 800 | 0.07°C 0.02% + 0.05°C | 0.14°C 0.04% + 0.10°C | |||||
| 200 Ω Pt (385) | -200 to 100 100 to 630 | 0.07°C 0.02% + 0.05°C | 0.14°C 0.04% + 0.10°C | |||||
| 500 Ω Pt (385) | -200 to 100 100 to 630 | 0.07°C 0.02% + 0.05°C | 0.14°C 0.04% + 0.10°C | |||||
| 1000 Ω Pt (385) | -200 to 100 100 to 630 | 0.07°C 0.02% + 0.05°C | 0.14°C 0.04% + 0.10°C | |||||
| 100 Ω Pt (3916) | -200 to 100 100 to 630 | 0.07°C 0.02% + 0.05°C | 0.14°C 0.04% + 0.10°C | |||||
| 100 Ω Pt (3926) | -200 to 100 100 to 630 | 0.08°C 0.02% + 0.06°C | 0.16°C 0.04% + 0.12°C | |||||
| 10 Ω Cu (427) | -100 to 260 | 0.2°C | 0.4°C | |||||
| 120 Ω Ni (672) | -80 to 260 | 0.1°C | 0.2°C | |||||
| Source current | Source °C | Allowable current2 | ||||||
| 1 year | 2 year | |||||||
| 1 mA | 0.05°C 0.0125% + 0.04°C | 0.10°C 0.025% + 0.08°C | 0.1 mA to 10 mA | |||||
| 500 μA | 0.06°C 0.017% + 0.05°C | 0.12°C 0.034% + 0.10°C | 0.1 mA to 1 mA | |||||
| 250 μA | 0.06°C 0.017% + 0.05°C | 0.12°C 0.034% + 0.10°C | 0.1 mA to 1 mA | |||||
| 150 μA | 0.06°C 0.017% + 0.05°C | 0.12°C 0.034% + 0.10°C | 0.1 mA to 1 mA | |||||
| 1 mA | 0.05°C 0.0125% + 0.04°C | 0.10°C 0.025% + 0.08°C | 0.1 mA to 10 mA | |||||
| 1 mA | 0.05°C 0.0125% + 0.04°C | 0.10°C 0.025% + 0.08°C | 0.1 mA to 10 mA | |||||
| 3 mA | 0.2°C | 0.4°C | 0.1 mA to 10 mA | |||||
| 1 mA | 0.04°C | 0.08°C | 0.1 mA to 10 mA | |||||
| 1. For two and three-wire RTD measurements, add 0.4°C to the specifications. 2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms | ||||||||
| Temperature, Thermocouples | ||||||||
| Type | Source °C | Measure °C | Source °C | |||||
| 1 year | 2 years | 1 year | 2 years | |||||
| E | -250 to -200 | 1.3 | 2.0 | 0.6 | 0.9 | |||
| -200 to -100 | 0.5 | 0.8 | 0.3 | 0.4 | ||||
| -100 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | ||||
| 600 to 1000 | 0.4 | 0.6 | 0.2 | 0.3 | ||||
| N | -200 to -100 | 1.0 | 1.5 |
Fluke 753 Documenting Process Calibrator
Tags: FLUKE
Official Facebook Pageเนื้อหาส่วนนี้อาจไม่แสดงกรณีใช้ผ่านมือถือบางรุ่น ทางบริษัทฯ ขออภัยในความไม่สะดวก แคตตาล็อก 2569บริษัท ฯ ไม่อนุญาตให้นำภาพไปใช้ในทุกกรณี | ||||