Range: 2 mV to 1000V 10 Hz to 1 MHz Best 1 Year Specification: ± 10ppm
Temperature Stabilization
Allow 12 hours stabilization time in the environment of use
Warm up Time
15 minutes with power on, after stabilization time
Temperature Performance
Operating: 11°C to 35°C; Calibration: 18°C to 28°C; Storage: -40°C to 50°C
Relative Humidity
Operating: < 75% to 30°C, < 70% to 35°C; Storage: < 95%, non-condensing
Altitude
Operating: to 3,050 m (10,000 ft); Non-operating: to 12,200 m (40,000 ft)
Input Low Isolation
20V to chassis
Guard Isolation
10V to input LO or chassis
EMI/RFI
Designed to comply with FCC Rules Part 15, Subpart J, Class B; VDE 0871, Class B; VDE 0875, Class K
Line Power
MIL-STD-28800D, paragraph 3.13.3
Calibration Documentation
Report of calibration with data included; 17025 accredited report or Z520.1 report optional
Size
Transfer Unit: 17.8 cm H x 21.6 cm W x 30.5 cm D (7 in H x 8.5 in W x 12 in D) Power Pack: 17.8 cm H x 21.6 cm W x 30.5 cm D (7 in H x 8.5 in W x 12 in D) 1000V Range Resistor: 7.6 cm H x 8.9 cm W x 14.0 cm D (3 in H x 3.5 in W x 5.5 in D) Transfer Switch: 7.6 cm H x 8.9 cm W x 14.0 cm D (3 in H x 3.5 in W x 5.5 in D)
Weight
Transfer Unit: 8.4 kg (18.5 lb) Power Pack: 8.9 kg (19.5 lb) 1000V Range Resistor: 1.6 kg (3.5 lb) Transfer Switch: 1.6 kg (3.5 lb)