Operating: 0 °C to °50 C Calibration: 15 °C to 35 °C Storage: -40 °C to 70 °C
Electromagnetic Compatibility (EMC)
International
IEC 61326-1: Controlled Electromagnetic Environment - CISPR 11: Group 1, Class A
Korea (KCC)
Class A Equipment (Industrial Broadcasting & Communication Equipment)
USA (FCC)
47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Surge
ANSI C62.41-1980, Category A
Reliability
MIL-T-2880D, paragraph 3.13.3
General Specifications
Dimensions
Height: 17.8 cm (7 in) standard rackmount + 1.5 cm (0.6 in) Width: 43.2 cm (17 in) Depth: 62 cm (24.8 in)
Maximum power requirements
5790B: 100 VA
Weight
5790B: 24 kg (53 lb) With wideband: 24.5 kg (54 lb)
Line power
50 Hz or 60 Hz; 100 V - 120 V, 220 V - 240 V 100V, 110 V, 115 V, 120 V, 200 V, 220 V, 230 V, 240 V
Safety
Complies with UL1244 and IEC 348-1976 and IEC 1010 and CSA C22.2 No. 231 and ANSI/ISA S82
Remote interfaces
RS-232, IEEE-488
Confidence level
99% unless otherwise specified
5790B/AF
The 5790B/AF has enhanced absolute accuracy to ± 0.23 % of voltage (1 year after factory calibration, 23 °C ± 3 °C, at 95 % confidence level) at 50 MHz and 223.61 mV, using a serialized one meter cable, characterized at that point for RF power meter 0 dBm reference output measurement. In the RF power measurement mode, the accuracy is specified to the end of the serialized cable with a curve fit for the other ranges and frequencies that is a first order correction for cable loss.