Fluke 1587 FC Insulation Multimeter
Specifications: Fluke 1587 FC Insulation Multimeter
Electrical Specifications AC voltage measurement Range 600.0 mV Resolution 0.1 mV Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) ±(1% + 3) Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) ±(2% + 3) 6.000 V Resolution 0.001 V Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) ±(1% + 3) Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) ±(2% + 3) 60.00 V Resolution 0.01 V Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) ±(1% + 3) Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) ±(2% + 3) 600.0 V Resolution 0.1 V Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) ±(1% + 3) Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) ±(2% + 3)¹ 1000 V Resolution 1 V Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) ±(2% + 3) Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) ±(2% + 3)¹ ¹ 1 kHz bandwidth Low-Pass Filter Voltage Range 600.0 mV Resolution 0.1 mV Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) ±(1% + 3) Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) +(2% + 3), -(6% - 3) 6.000 V Resolution 0.001 V Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) ±(1% + 3) Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) +(2% + 3), -(6% - 3) 60.00 V Resolution 0.01 V Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) ±(1% + 3) Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) +(2% + 3), -(6% - 3) 600.0 V Resolution 0.1 V Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) ±(1% + 3) Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) +(2% + 3), -(6% - 3) 1000 V Resolution 1 V Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) ±(2% + 3) Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) +(2% + 3), -(6% - 3) DC Voltage Measurement Range Resolution Accuracy ±(% of Rdg + Counts) 6.000 V dc 0.001 V ±(0.09% + 2) 60.00 V dc 0.01 V ±(0.09% + 2) 600.0 V dc 0.1 V ±(0.09% + 2) 1000 V dc 1 V ±(0.09% + 2) Input impedance 10 MΩ (nominal), <100 pF Normal mode rejection ratio >60 dB @ 50 Hz or 60 Hz Common mode rejection ratio >120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance) Accuracies apply to ± 100% of range DC Millivolts Measurement Range Resolution Accuracy ±(% of Rdg + Counts) 600.0 mVdc 0.1 mV ±(0.1% + 1) DC and AC Current Measurement AC 45 Hz to 1000 Hz Range 400 mA Resolution 0.1 mA Accuracy ±(% of Rdg+Counts) ±(1.5% + 2)¹ Burden voltage (Typical) 2 mV/mA Range 60 mA Resolution 0.01 mA Accuracy ±(% of Rdg+Counts) ±(1.5% + 2)¹ Burden voltage (Typical) 2 mV/mA DC Range 400 mA Resolution 0.1 mA Accuracy ±(% of Rdg+Counts) ±(0.2% + 2) Burden voltage (Typical) 2 mV/mA Range 60 mA Resolution 0.01 mA Accuracy ±(% of Rdg+Counts) ±(0.2% + 2) Burden voltage (Typical) 2 mV/mA Overload 600 mA for 2 minutes maximum Fuse protection for mA input 0.44 mA, 1000 V, IR 10 kA AC conversion Inputs are ac-coupled and calibrated to the rms value of sine wave input Conversions are true-rms responding and specified from 5% to 100% of range. Input signal crest factor can be up to 3 up to 300 mA, decreasing linearly to crest factor ≤1.5 at 600 mA. For non-sinusoidal waveforms add +(2% reading + 2% FS) typical, for a crest factor up to 3. ¹ 1 kHz bandwidth Ohms Measurement Range Resolution Accuracy +(% of Rdg+Counts)¹ 600.0 Ω 0.1 Ω ±(0.9% + 2) 6.000 kΩ 0.001 kΩ ±(0.9% + 2) 60.00 kΩ 0.01 kΩ ±(0.9% + 2) 600.0 kΩ 0.1 KΩ ±(0.9% + 2) 6.000 MΩ 0.001 MΩ ±(0.9% + 2) 50.0 MΩ ² 0.01 MΩ ±(1.5% + 3) Overload protection 1000 V rms or dc Open circuit test voltage <8.0 V dc Short circuit current <1.1 mA ¹ Accuracies apply from 0% to 100% of range
² Up to 80% relative humidity Diode Test Diode test indication Display voltage drop: 0.6 V at1.0 mA nominal test current: Accuracy ±(2% + 3) Continuity Test Continuity indication Continuous audible tone for testresistance below 25 Ω and off above 100 Ω. Maximum Reading; 1000 Ω Open circuit voltage <8.0 V Short circuit current 1.0 mA typical Overload protection 1000 V rms Response time >1 m sec Frequency Measurement Range Resolution Accuracy ±(% of Rdg+Counts) 99.99 Hz 0.01 Hz ±(0.1% + 1) 999.9 Hz 0.1 Hz ±(0.1% + 1) 9.999 kHz 0.001 kHz ±(0.1% + 1) 99.99 kHz 0.01 kHz ±(0.1% + 1) Frequency Counter Sensitivity 600.0 mV ac V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz 100.0 mV V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz 150.0 mV DC Trigger Levels¹ to 20 kHz² NA 6.0 V V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz 1.0 V V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz 1.5 V DC Trigger Levels¹ to 20 kHz² -400.0 mV and 2.5 V 60.0 V V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz 10.0 V V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz 36.0 V DC Trigger Levels¹ to 20 kHz² 1.2 V and 4.0 V 600.0 V V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz 100.0 V V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz - DC Trigger Levels¹ to 20 kHz² 12.0 V and 40.0 V 1000.0 V V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz 300.0 V V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz - DC Trigger Levels¹ to 20 kHz² 12.0 V and 40.0 V ¹ Maximum input for specified accuracy = 10x range (1000 V max). Noise at low frequencies and amplitudes may affect accuracy
² Usable to 100 kHz with full scale input Capacitance Range Resolution Accuracy ±(% of Rdg+Counts) 1000 nF 1 nF ±(1.2% + 2) 10.00 μF 0.01 μF ±(1.2% + 2) 100.0 μF 0.1 μF ±(1.2% ±90 counts) 9999 μF 1 μF ±(1.2% ±90 counts) Temperature Measurement Range Resolution Accuracy ±(% of Rdg+Counts)¹ -40 ° C to 537 ° C 0.1 °C ±(1% + 10 counts) -40 ° F to 998 ° F 0.1 °F ±(1% + 18 counts) ¹ Accuracies apply following 90 minutes settling time after a change in the ambient temperature of the instrument Insulation Specifications Measurement range 0.01 MΩ to 2 GΩ Test voltages 50, 100, 250, 500, 1000 V Test voltage accuracy +20%, -0% Short-circuit test current 1 mA nominal Auto discharge Discharge time <0.5 second for C = 1 μF or less Live circuit detection Inhibit test if terminal voltage > 30 V prior to initialization of test Maximum capacitive load Operable with up to 1 μF load Output Voltage 50 V (0% to +20%) Display range
Fluke 1587 FC Insulation Multimeter
- Product Code:Fluke 1587 FC Insulation Multimeter
- Availability:In Stock
Tags: FLUKE

สินค้าของเรา
เครื่องกำเนิดไฟฟ้า ·
รอกยกของ · ไม้ชักฟิวส์ · ไม้ฉนวน · บันไดสไลด์ · มีดปอกสาย
· ชุดประกอบฉนวนค้ำยันสายไฟ · Ratchet Hoist แบบผ้าใบ · Ratchet Cable Cutter · Insulator Cover · InsulatedHanger
· Conductor Cutter with Insulated Handles · Conductor Cover 46 kv · Clamp stick grip
all type ขนาด 8 ฟุต · Clamp Stick
grip all type ขนาด 12 ฟุต · สว่านไร้สาย · ล้อวัดระยะทาง · ประแจบล๊อคไฟฟ้าไร้สาย (BRUSHLESS CORDLESS IMPACT WRENCH) · เทปวัดระยะชนิดโลหะ · ตู้แผงสวิตช์ · ตู้เก็บเครื่องมือ · เครื่องอัดลมใช้มอเตอร์ไฟฟ้า ·
เครื่องมือบีบหลอดต่อสายไฮดรอลิค · เครื่องมือบีบคอนเนคเตอร์
แบบ H · เครื่องมือต่อสายลงดิน (Grounding Equipment)
· เครื่องมือจับและดึงสาย (Wire Grip Come Along) Overall
outside · เครื่องดึงสาย (Coffing Hoist) · อุปกรณ์ระบบไฟฟ้า
· สวิตซ์ตัดตอนแรงสูง · หางปลา · หลอดต่อสาย · สลักเข้าปลายสาย · พีจีคอนเนคเตอร์ · ฉนวนป้องกันไฟฟ้าลัดวงจร ·
ฉนวนครอบล่อฟ้า · เครื่องมือช่างไฟฟ้าอุตสาหกรรม · เครื่องมือไฮดรอลิก · เครื่องมือแบตเตอรี่ไร้สายดิสคอนเนคติ้งสวิตซ์แบริเออร์
· ฉนวนครอบบุชชิ่ง · ตาข่าย · แผ่นอลูมิเนียม · ฉนวนครอบดรอปเอาท์ฟิวส์ · หนามป้องกันสัตว์ · อุปกรณ์ระบบไฟฟ้า · หม้อแปรงไฟฟ้า · ROADMAPTOSTRONGGRIDLV · แคล้มป์สติ๊ก
· ClampStick · เชือกใยยักษ์ · เชือก Handline
· KNIPEX · เครื่องตรวจจับแรงดันไฟฟ้า · UNI · สายกันตก · UNI9103 · เข็มขัดปีนเสา · YQK ·
SME · ขายดี · หมวกปีกรอบ · หมวกนิรภัย S-Guard · MSA V-Gard full brim · คีมตัดสาย · คัทเตอร์ตัดสาย · CableCutter · คีมตัดลวดเหล็ก ·
คัทเตอร์เฟืองทด · คัทเตอร์ตัดสายเคเบิ้ล ·
คีมย้ำ · คีมบีบ · แคล้มป์สติ๊ก
·
KANDP · Milwaukee · CHANCE · RITZ · KUDOS · ZUPPER · FLUKE · Klauke Hydraulic Crimp Tool · Google ·
Official Facebook Page
เนื้อหาส่วนนี้อาจไม่แสดงกรณีใช้ผ่านมือถือบางรุ่น ทางบริษัทฯ ขออภัยในความไม่สะดวก
4 TOP Facebook Official Page Post
เนื้อหาส่วนนี้อาจไม่แสดงกรณีใช้ผ่านมือถือบางรุ่น ทางบริษัทฯ ขออภัยในความไม่สะดวก
Electrical Specifications | ||||||||||
AC voltage measurement | ||||||||||
Range | ||||||||||
600.0 mV | Resolution | 0.1 mV | ||||||||
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(1% + 3) | |||||||||
Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) | ±(2% + 3) | |||||||||
6.000 V | Resolution | 0.001 V | ||||||||
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(1% + 3) | |||||||||
Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) | ±(2% + 3) | |||||||||
60.00 V | Resolution | 0.01 V | ||||||||
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(1% + 3) | |||||||||
Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) | ±(2% + 3) | |||||||||
600.0 V | Resolution | 0.1 V | ||||||||
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(1% + 3) | |||||||||
Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) | ±(2% + 3)¹ | |||||||||
1000 V | Resolution | 1 V | ||||||||
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(2% + 3) | |||||||||
Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) | ±(2% + 3)¹ | |||||||||
¹ 1 kHz bandwidth | ||||||||||
Low-Pass Filter Voltage | ||||||||||
Range | ||||||||||
600.0 mV | Resolution | 0.1 mV | ||||||||
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(1% + 3) | |||||||||
Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) | +(2% + 3), -(6% - 3) | |||||||||
6.000 V | Resolution | 0.001 V | ||||||||
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(1% + 3) | |||||||||
Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) | +(2% + 3), -(6% - 3) | |||||||||
60.00 V | Resolution | 0.01 V | ||||||||
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(1% + 3) | |||||||||
Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) | +(2% + 3), -(6% - 3) | |||||||||
600.0 V | Resolution | 0.1 V | ||||||||
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(1% + 3) | |||||||||
Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) | +(2% + 3), -(6% - 3) | |||||||||
1000 V | Resolution | 1 V | ||||||||
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(2% + 3) | |||||||||
Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) | +(2% + 3), -(6% - 3) | |||||||||
DC Voltage Measurement | ||||||||||
Range | Resolution | Accuracy ±(% of Rdg + Counts) | ||||||||
6.000 V dc | 0.001 V | ±(0.09% + 2) | ||||||||
60.00 V dc | 0.01 V | ±(0.09% + 2) | ||||||||
600.0 V dc | 0.1 V | ±(0.09% + 2) | ||||||||
1000 V dc | 1 V | ±(0.09% + 2) | ||||||||
Input impedance | 10 MΩ (nominal), <100 pF | |||||||||
Normal mode rejection ratio | >60 dB @ 50 Hz or 60 Hz | |||||||||
Common mode rejection ratio | >120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance) | |||||||||
Accuracies apply to ± 100% of range | ||||||||||
DC Millivolts Measurement | ||||||||||
Range | Resolution | Accuracy ±(% of Rdg + Counts) | ||||||||
600.0 mVdc | 0.1 mV | ±(0.1% + 1) | ||||||||
DC and AC Current Measurement | ||||||||||
AC 45 Hz to 1000 Hz | ||||||||||
Range | 400 mA | |||||||||
Resolution | 0.1 mA | |||||||||
Accuracy ±(% of Rdg+Counts) | ±(1.5% + 2)¹ | |||||||||
Burden voltage (Typical) | 2 mV/mA | |||||||||
Range | 60 mA | |||||||||
Resolution | 0.01 mA | |||||||||
Accuracy ±(% of Rdg+Counts) | ±(1.5% + 2)¹ | |||||||||
Burden voltage (Typical) | 2 mV/mA | |||||||||
DC | ||||||||||
Range | 400 mA | |||||||||
Resolution | 0.1 mA | |||||||||
Accuracy ±(% of Rdg+Counts) | ±(0.2% + 2) | |||||||||
Burden voltage (Typical) | 2 mV/mA | |||||||||
Range | 60 mA | |||||||||
Resolution | 0.01 mA | |||||||||
Accuracy ±(% of Rdg+Counts) | ±(0.2% + 2) | |||||||||
Burden voltage (Typical) | 2 mV/mA | |||||||||
Overload | 600 mA for 2 minutes maximum | |||||||||
Fuse protection for mA input | 0.44 mA, 1000 V, IR 10 kA | |||||||||
AC conversion | Inputs are ac-coupled and calibrated to the rms value of sine wave input | |||||||||
Conversions are true-rms responding and specified from 5% to 100% of range. Input signal crest factor can be up to 3 up to 300 mA, decreasing linearly to crest factor ≤1.5 at 600 mA. For non-sinusoidal waveforms add +(2% reading + 2% FS) typical, for a crest factor up to 3. | ||||||||||
¹ 1 kHz bandwidth | ||||||||||
Ohms Measurement | ||||||||||
Range | Resolution | Accuracy +(% of Rdg+Counts)¹ | ||||||||
600.0 Ω | 0.1 Ω | ±(0.9% + 2) | ||||||||
6.000 kΩ | 0.001 kΩ | ±(0.9% + 2) | ||||||||
60.00 kΩ | 0.01 kΩ | ±(0.9% + 2) | ||||||||
600.0 kΩ | 0.1 KΩ | ±(0.9% + 2) | ||||||||
6.000 MΩ | 0.001 MΩ | ±(0.9% + 2) | ||||||||
50.0 MΩ ² | 0.01 MΩ | ±(1.5% + 3) | ||||||||
Overload protection | 1000 V rms or dc | |||||||||
Open circuit test voltage | <8.0 V dc | |||||||||
Short circuit current | <1.1 mA | |||||||||
¹ Accuracies apply from 0% to 100% of range ² Up to 80% relative humidity | ||||||||||
Diode Test | ||||||||||
Diode test indication | Display voltage drop: 0.6 V at1.0 mA nominal test current: | |||||||||
Accuracy | ±(2% + 3) | |||||||||
Continuity Test | ||||||||||
Continuity indication | Continuous audible tone for testresistance below 25 Ω and off above 100 Ω. Maximum Reading; 1000 Ω | |||||||||
Open circuit voltage | <8.0 V | |||||||||
Short circuit current | 1.0 mA typical | |||||||||
Overload protection | 1000 V rms | |||||||||
Response time | >1 m sec | |||||||||
Frequency Measurement | ||||||||||
Range | Resolution | Accuracy ±(% of Rdg+Counts) | ||||||||
99.99 Hz | 0.01 Hz | ±(0.1% + 1) | ||||||||
999.9 Hz | 0.1 Hz | ±(0.1% + 1) | ||||||||
9.999 kHz | 0.001 kHz | ±(0.1% + 1) | ||||||||
99.99 kHz | 0.01 kHz | ±(0.1% + 1) | ||||||||
Frequency Counter Sensitivity | ||||||||||
600.0 mV ac | V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz | 100.0 mV | ||||||||
V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz | 150.0 mV | |||||||||
DC Trigger Levels¹ to 20 kHz² | NA | |||||||||
6.0 V | V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz | 1.0 V | ||||||||
V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz | 1.5 V | |||||||||
DC Trigger Levels¹ to 20 kHz² | -400.0 mV and 2.5 V | |||||||||
60.0 V | V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz | 10.0 V | ||||||||
V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz | 36.0 V | |||||||||
DC Trigger Levels¹ to 20 kHz² | 1.2 V and 4.0 V | |||||||||
600.0 V | V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz | 100.0 V | ||||||||
V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz | - | |||||||||
DC Trigger Levels¹ to 20 kHz² | 12.0 V and 40.0 V | |||||||||
1000.0 V | V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz | 300.0 V | ||||||||
V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz | - | |||||||||
DC Trigger Levels¹ to 20 kHz² | 12.0 V and 40.0 V | |||||||||
¹ Maximum input for specified accuracy = 10x range (1000 V max). Noise at low frequencies and amplitudes may affect accuracy ² Usable to 100 kHz with full scale input | ||||||||||
Capacitance | ||||||||||
Range | Resolution | Accuracy ±(% of Rdg+Counts) | ||||||||
1000 nF | 1 nF | ±(1.2% + 2) | ||||||||
10.00 μF | 0.01 μF | ±(1.2% + 2) | ||||||||
100.0 μF | 0.1 μF | ±(1.2% ±90 counts) | ||||||||
9999 μF | 1 μF | ±(1.2% ±90 counts) | ||||||||
Temperature Measurement | ||||||||||
Range | Resolution | Accuracy ±(% of Rdg+Counts)¹ | ||||||||
-40 ° C to 537 ° C | 0.1 °C | ±(1% + 10 counts) | ||||||||
-40 ° F to 998 ° F | 0.1 °F | ±(1% + 18 counts) | ||||||||
¹ Accuracies apply following 90 minutes settling time after a change in the ambient temperature of the instrument | ||||||||||
Insulation Specifications | ||||||||||
Measurement range | 0.01 MΩ to 2 GΩ | |||||||||
Test voltages | 50, 100, 250, 500, 1000 V | |||||||||
Test voltage accuracy | +20%, -0% | |||||||||
Short-circuit test current | 1 mA nominal | |||||||||
Auto discharge | Discharge time <0.5 second for C = 1 μF or less | |||||||||
Live circuit detection | Inhibit test if terminal voltage > 30 V prior to initialization of test | |||||||||
Maximum capacitive load | Operable with up to 1 μF load | |||||||||
Output Voltage | ||||||||||
50 V (0% to +20%) | Display range |
Fluke 1587 FC Insulation Multimeter
Tags: FLUKE สินค้าของเราเครื่องกำเนิดไฟฟ้า ·
รอกยกของ · ไม้ชักฟิวส์ · ไม้ฉนวน · บันไดสไลด์ · มีดปอกสาย
· ชุดประกอบฉนวนค้ำยันสายไฟ · Ratchet Hoist แบบผ้าใบ · Ratchet Cable Cutter · Insulator Cover · InsulatedHanger
· Conductor Cutter with Insulated Handles · Conductor Cover 46 kv · Clamp stick grip
all type ขนาด 8 ฟุต · Clamp Stick
grip all type ขนาด 12 ฟุต · สว่านไร้สาย · ล้อวัดระยะทาง · ประแจบล๊อคไฟฟ้าไร้สาย (BRUSHLESS CORDLESS IMPACT WRENCH) · เทปวัดระยะชนิดโลหะ · ตู้แผงสวิตช์ · ตู้เก็บเครื่องมือ · เครื่องอัดลมใช้มอเตอร์ไฟฟ้า ·
เครื่องมือบีบหลอดต่อสายไฮดรอลิค · เครื่องมือบีบคอนเนคเตอร์
แบบ H · เครื่องมือต่อสายลงดิน (Grounding Equipment)
· เครื่องมือจับและดึงสาย (Wire Grip Come Along) Overall
outside · เครื่องดึงสาย (Coffing Hoist) · อุปกรณ์ระบบไฟฟ้า
· สวิตซ์ตัดตอนแรงสูง · หางปลา · หลอดต่อสาย · สลักเข้าปลายสาย · พีจีคอนเนคเตอร์ · ฉนวนป้องกันไฟฟ้าลัดวงจร ·
ฉนวนครอบล่อฟ้า · เครื่องมือช่างไฟฟ้าอุตสาหกรรม · เครื่องมือไฮดรอลิก · เครื่องมือแบตเตอรี่ไร้สายดิสคอนเนคติ้งสวิตซ์แบริเออร์
· ฉนวนครอบบุชชิ่ง · ตาข่าย · แผ่นอลูมิเนียม · ฉนวนครอบดรอปเอาท์ฟิวส์ · หนามป้องกันสัตว์ · อุปกรณ์ระบบไฟฟ้า · หม้อแปรงไฟฟ้า · ROADMAPTOSTRONGGRIDLV · แคล้มป์สติ๊ก
· ClampStick · เชือกใยยักษ์ · เชือก Handline
· KNIPEX · เครื่องตรวจจับแรงดันไฟฟ้า · UNI · สายกันตก · UNI9103 · เข็มขัดปีนเสา · YQK ·
SME · ขายดี · หมวกปีกรอบ · หมวกนิรภัย S-Guard · MSA V-Gard full brim · คีมตัดสาย · คัทเตอร์ตัดสาย · CableCutter · คีมตัดลวดเหล็ก ·
คัทเตอร์เฟืองทด · คัทเตอร์ตัดสายเคเบิ้ล ·
คีมย้ำ · คีมบีบ · แคล้มป์สติ๊ก
·
KANDP
Official Facebook Pageเนื้อหาส่วนนี้อาจไม่แสดงกรณีใช้ผ่านมือถือบางรุ่น ทางบริษัทฯ ขออภัยในความไม่สะดวก 4 TOP Facebook Official Page Post
เนื้อหาส่วนนี้อาจไม่แสดงกรณีใช้ผ่านมือถือบางรุ่น ทางบริษัทฯ ขออภัยในความไม่สะดวก |